Description
Sensitivity | |||
Sensor Type | 90° light scattering, 650 nm laser diode | ||
Calibration | Calibrated against a gravimetric reference using the respirable fraction of standard ISO 12103-1, A1 Test Dust | ||
Aerosol Concentration Range | 0.001 to 100 mg/m3 | ||
Particle Size Range | 0.1 to 10 μm | ||
Minimum Resolution | 0.001 mg/m3 | ||
Zero Stability | ±0.001 mg/m3 over 24 hours using 10 second time constant | ||
Temperature Coefficient | Approximately +0.0005 mg/m3 per °C (for variations from temperature at which instrument was last zeroed) | ||
Flow Rate | |||
Range | User-adjustable, 0 to 1.8 liters/min (L/min) Note: Upper range is dependent on impactor or cyclone attached to the SidePak AM520 Monitor | ||
Temperature Range | |||
Operating Range | 32° F to 120° F (0° C to 50° C) | ||
Storage Range | -4° F to 140° F (-20° C to 60° C) | ||
Operational Humidity | 0 to 95% RH, non-condensing | ||
Time Constant (OLED Display) | |||
Range | User-adjustable, from 1 to 60 seconds | ||
Built-in Inlets | |||
Standard inlet | Not size-specific | ||
PM1.0 impactor | 50% cut-off at 1.0 μm | ||
PM2.5 impactor | 50% cut-off at 2.5 μm | ||
PM5.0 impactor | 50% cut-off at 5.0 μm | ||
PM10 impactor | 50% cut-off at 10.0 μm | ||
Attachable Cyclones | |||
4 μm Dorr-Oliver | 50% cut-off at 4.0 μm | ||
0.8 μm DPM | 50% cut-off at 0.8 μm | ||
Alarms | |||
Alarm Types | General, STEL | ||
Default Settings | 0.5 mg/m3, 1 mg/m3, 3 mg/m3, 5 mg/m3, 15 mg/m3, Off | ||
Programmable from 0.001 to 999 mg/m3 through TrakPro Software | |||
Alarm Indicator | 95 dBA (at one foot) audible horn, blinking red LED, flashing red OLED display | ||
Ringback Delay | 30 sec, 1 min, 3 min, 5 min, 10 min, Off | ||
Data Logging | |||
Data Points | Approx. 80,000 (55 days logging once per minute) | ||
Logging Interval | User-adjustable, from 1 second to 60 minutes | ||
User-Select Calibration Factors | |||
Factory Setting | 1.0 (non-adjustable) | ||
Ambient Setting | 0.38 (non-adjustable) | ||
User-Defined Settings | 5, with user-defined labels via TrakPro Software | ||
Range | 0.1 to 10.0, user-adjustable | ||
Physical | |||
External Dimensions | 5.1 x 3.7 x 3.1 inch129.5 mm x 94 mm x 78.4 mm with 803300 battery | ||
Weight | 21 oz (0.62 kg) with 803300 battery | ||
Display | 159 x 128 resolution color OLED display | ||
Tripod Mounting Nut | ¼-20 female thread | ||
Battery | |||
Rating | 5399 mAH Li-Ion Pack | ||
Run Time | Greater than 20 hours at 1.7Lpm with a PM2.5 impactor | ||
Charge Time | 4.0 hours (typical with a fully depleted battery) | ||
Power Supply (p/n 803302) | |||
Input Voltage Range | 100 to 240 VAC, 50 to 60 Hz | ||
Output Voltage | 12 VDC @ 3.0A | ||
Maintenance | |||
Factory Clean/Calibrate | Recommended annually | ||
User Zero Calibration | Before each use | ||
User Flow Calibration | As needed with use of inlet conditioner | ||
Communications Interface | |||
Type | USB 2.0 | ||
Connector, Instrument | USB Micro-B (socket) | ||
Minimum Computer Requirements for TrakPro Software | |||
Communications Port | Universal Serial Bus (USB) v2.0 or higher | ||
Operating System | Microsoft Windows® 7, 8, or 10 (32-bit or 64-bit) operating systems | ||
Approvals | |||
AM520 with TSI Battery Pack P/N 803300 | |||
Immunity | EN61326-1:2013 | ||
Emissions | EN61326-1:2013 Class B | ||
Safety | IEC 61010-1:2010 | ||
IEC | 60825-1:2014 |